Opened 9 years ago

Closed 9 years ago

#265 closed defect (fixed)

Frontend 2DSA band_forming parameters

Reported by: gegorbet Owned by: gegorbet
Priority: major Milestone: future
Component: ultrascan3 Version:
Keywords: review Cc:

Description (last modified by demeler)

A couple of additional points:

  • the volume should have limits of 1-20 microliters, with a default of 15
  • the simulation points should be from 50-10,000, with a default of 200
  • simulation points counter should have 3 buttons (10/100/1000)

Change History (3)

comment:1 in reply to: ↑ description Changed 9 years ago by demeler

  • Description modified (diff)

Replying to gegorbet:

Borries reported and I was able to verify that there are some problems with advanced parameters, particularly related to band-forming centerpiece, with frontend 2DSA. Some of these issues also exist for us_astfem_sim. They are the following.

  • The band-forming flag is not automatically set from the centerpiece sector value.
  • In advanced settings, setting band-forming does not immediately enable the band_volume counter.
  • The band_volume input is in milliliters; it should be in microliters.
  • The band_volume range and default are wrong.
  • The Simulation Points input has wrong range and default.

comment:2 Changed 9 years ago by gegorbet

  • Keywords review added
  • Status changed from new to assigned

The GUI changes to both 2DSA and astfem_sim have been made, so that variables related to band forming have correct ranges and defaults. In addition, the astfem_rsa utility has been modified to pick up angle and pathlength values in its calculation of lamella width. This last affects both frontend and backend 2DSA.

Review-ready.

comment:3 Changed 9 years ago by gegorbet

  • Resolution set to fixed
  • Status changed from assigned to closed
Note: See TracTickets for help on using tickets.