  • section-detectionsystems/detectionsystems.tex

    r1 r5  
    1717\caption{Emission spectra for a Xenon-flash lamp (electric discharge) with oil deposited onto the light bulb (graph A) and after lamp cleaning with a cotton tip swab, toothpaste, water and ethanol (graph B).}
    1818\label{fig-esxfl}
    19 \end{center}\end{figure}\\ \indent After turning off the power, the Xenon flash lamp assembly can be removed noting the orientation of the lamp assembly by marking it with tape or ink. The cover is removed and the lamp cleaned with toothpaste and a cotton tip swab or a pencil eraser (Figure \ref{fig-cxfl}). The lamp is cleaned further by ethanol and milli-Q water, and reassembled. The lamp is replaced in the same orientation as it was removed. Another diagnostic wavelength intensity scan can be performed for confirmation. Lamp cleaning is essential for a high signal-to-noise ratio to obtain quality data and can, in principle, be performed by the user. A trained service engineer or lab technician can guide a new user through the procedure.\begin{figure}[htbp]
     19\end{center}\end{figure}\\ \indent After turning off the power, the Xenon flash lamp assembly can be removed noting the orientation of the lamp assembly by marking it with tape or ink. The cover is removed and the lamp cleaned with toothpaste and a cotton tip swab or a pencil eraser (Figure \ref{fig-cxfl}). The lamp is cleaned further by ethanol and Milli-Q water, and reassembled. The lamp is replaced in the same orientation as it was removed. Another diagnostic wavelength intensity scan can be performed for confirmation. Lamp cleaning is essential for a high signal-to-noise ratio to obtain quality data and can, in principle, be performed by the user. A trained service engineer or lab technician can guide a new user through the procedure.\begin{figure}[htbp]
    2020\begin{center}
    2121\includegraphics[width=0.9\linewidth]{section-detectionsystems/CareOfTheXenonLamp.png}